GDZT-50-200-40

Autonomous products and other conventional products


Classification:

-40-200 ℃ high and low temperature circulation device


Key words: experimental instrument


online message

Product Details

Scientific instruments are tools for precision measurement, and their precision and reliability must be an order of magnitude higher than the object being measured. Therefore, although there are not as many components as airplanes, the systematic design is also beyond imagination. The chromatographic column, as a consumable, is a separation center, just like the heart. High pressure, tightness and flow rate stability are required. And the oven needs to be kept at a constant temperature. To prevent leakage, maintain constant temperature, pressurized flow, every link, can not be slack. For materials, seals, and fluid mechanics, a high foundation is required.

The method for checking whether the instrument leaks: pinch off the external vacuum skin tube, and observe whether the vacuum gauge on the instrument can keep the instrument leak-free for five minutes. If there is air leakage, check whether the sealing ring on each sealing joint and rotating shaft is effective. On the contrary, the instrument is normal to check the vacuum pump and vacuum pipeline.

Working principle

This product has both cooling function and heating function. It is composed of refrigeration unit, heating unit, circulation system and control system.

The refrigeration part is the use of fully enclosed compressor mechanical refrigeration, air-cooled condenser, plate heat exchanger, expansion valve, pressure protector and other components to meet the cooling requirements; heating part is the use of electric heating tube to heat the circulating liquid. The refrigerant or heat medium is transported to the interlayer of the supporting equipment through the built-in circulating pump and the external circulating pipeline to indirectly cool down and heat the materials in the reactor.

During external circulation, the upper outlet pipeline of the product is connected to the lower inlet pipeline of the kettle body, and the circulating liquid returns to the lower outlet of the product through the pipeline from the upper circulation port to form a complete circulation space.

Technical parameters

Name

High and low temperature circulation device

Model

GDZT-50-200-40

Use temperature range

-40 ~ 200

Ambient temperature

5~35 ℃

Ambient humidity

≤ 70% ventilation

Power

Three-phase 380V/50HZ

Security Protection

Delay, overcurrent, overheating, overtemperature, phase sequence, phase loss protection

Display

LCD display, key operation

Temperature control accuracy

± 0.1 ℃

Sensor

PT100

Total power of whole machine

11395W

Heating power

6000W

Compressor

Specifications

6P

Quantity

1

Power

5145W

refrigerating capacity

16268W

Circulating pump

Power

250W

lift

15M

Flow

30L/min

Pressure

≤ 0.2MPa

plate heat exchanger

heat exchange area

1.46 ㎡

Refrigerant

R22

Body Material

Cold plate spray anti-corrosion

External Loop Interface

6 points wire mouth with 6 points × 2 meters of metal circulating pipe two

Circulating liquid

High and low temperature special heat transfer oil

Space size of machine

695 × 720 × 1100mm width depth height

machine net weight

156KG

Outer packing size

800 × 740 × 1270mm wooden case 0.76m ³

Packing weight

182KG